ICU Medical Plum Duo Precision Infusion Pump
Call For Price
Vendor direct - Longer lead times may apply
SKU
06-54-0002
Teach nursing students to deliver infusions on a modern system used in many hospitals. The Plum Duo™ IV Precision IV Pump provides dual-channel clinician-driven infusion technology that combines ease of use with proven medication delivery and safety.
Features:
- Unlike traditional pumps, Plum Duo provides unmatched ±3% accuracy, no matter where the bag, pump, or patient is positioned
- Designed with nurses in mind, Plum Duo simplifies workflows, leading to fewer burdens, smarter guidance, and more focused care
- Modern, intuitive interface:
- Vibrant high-definition durable display provides clear, critical information at a glance, reducing menu navigation
- Guides users through every step for faster programming and fewer errors
- Clear dose limits and automated calculations reduce errors, save time, and simplify care
- Provides a calmer environment:
- Plum Duo simplifies alarm handling with clear on-screen prompts, multiple prioritized alarms, and distal occlusion auto-reset
- Whisper-quiet motor enables a more peaceful environment
- Program pumps directly from patient orders and automacally document infusions using seamless IV-EHR interoperability
- Real-world precision:
- Plum Duo does not depend on gravity or perfect setups, consistently delivering ±3% accuracy
- Alerts nurses to closed clamps or obstructions, helping prevent errors and ensuring secondary medications are administered as intended
- Can infuse piggybacks to empty and flush the line
- Confidently manage complex infusions:
- Hang infusions where they make sense – no bag height concerns or complex setups; secondary infusions can be delivered directly from a syringe for flexibility
- Plum Duo delivers up to four compatible medications through one pump, reducing clutter and conserving space
- LifeShield Infusion Safety Software:
- Connects, informs, and empowers care teams to help deliver better patient care
- Software connects pharmacy, nursing, biomed, and IT to see the same data, share updates in near real-time, and keep IV medication management moving
- Keeps protocols sharp, helps reduce errors, and enhances responsiveness
- Includes drug library management, device manager, clinical dashboards and reports, and account manager for system-wide control from anywhere
- Clear and consistent drug library change summaries ensure clarity, offering a concise overview of updates; Pharmacy pushes protocols fleet-wide instantly, helping equip nurses with the latest safety standards to eliminate inconsistencies
Specifications:
Physical:
- Dimensions: Approx. 9" × 11.75" × 6.5" (excluding pole clamp extrusion and power cord storage)
- Two channels: Left and right channel. Each channel has primary (Line 1) and secondary (Line 2)
- Weight: Approx. 10.6 lb. with battery
- High-impact plastic casing
Electrical:
- Power Requirements: 100-240 VAC; 50-60 Hz; 160 VA
- Power Cord: Hospital-grade AC cord
- Battery: Lithium iron phosphate; 12.8V; 3,000 mAh; internal; rechargeable
- Battery Operation: Typical battery operating time with new and fully charged battery (nominal screen brightness) for one channel at 25 mL/hr. is 6:35 hr. and at 999 mL/hr. is 4:25 hours
- Both channels infusing at 25 mL/hr is 5:30 hr. and at 999 mL/hr. is 3 hours
- Recharge: Battery charges whenever the infusion pump is connected to AC (mains) power and it is below its maximum capacity
- Recharge time: Up to 8 hours
Environment:
- Operating: 41° F to 104° F (5° C to 40° C)
- Transporting and Storage: –4° F to 140° F (–20° C to 60° C)
- Atmospheric Pressure 0 to 10,000 ft (0 to 3,000 m) or equivalent atmospheric pressure
- Fluid Ingress Rating IPX2: Protected against vertical falling water drops
Delivery Rate, Range, and Accuracy:
- Delivery Accuracy: +/-3%: 0.1 to 999 mL/hr.; delivery accuracy testing was performed in accordance with IEC 60601-2-24:2012
- Lines 1 and 2: 0.1 to 9.99 mL/hr. (in 0.01 mL/hr. increments); 10 to 99.9 mL/hr. (in 0.1 mL/hr. increments); 100 to 999 mL/hr. (in 1 mL/hr. increments)
- Concurrent Delivery 0.5 mL/h minimum for each line; 500 mL/h cumulative (Line 1 + Line 2 on the same side) maximum
- KVO 1.0 mL/hr or the last primary delivery rate, whichever is less; NOTE: KVO may be configured by CCA to have a default setting of 1-20 mL/hr
- VTBI Range 0.1 to 99.9 mL (in 0.1 mL increments) and 100 to 9,999 mL (in 1 mL increments)
Occlusion Alarms and Limits:
- Downstream Occlusion: The downstream occlusion alarm sounds after the downstream tubing or set outlet fitting becomes occluded or vacuum occurs
- Upstream Occlusion: The upstream occlusion alarm sounds if the upstream to the cassette becomes occluded or pressurized
- Downstream Pressure Limit (Without Alarm): Maximum pressure limit: User-selectable factory default setting 6 psi (310 mm Hg); Selectable range: 1 to 15 psi (52 to 776 mm Hg) with display accuracy ± 3 psi (±155 mm Hg)
- Maximum Infusion Pressure 20 psi (1034 mm Hg)
Air-in-Line Alarms:
- PlumSet™ (Downstream) Air Bolus: Configured for each CCA as 50, 100, or 250 mL of air or larger; Cumulative: 0.25 mL of air out of 4.9 mL of fluid
- PlumSet (Upstream) Air Bolus: 0.5 mL of air or larger; Cumulative: 1 mL of air (0.5 mL of air per line for concurrent delivery)
Contact Pocket Nurse® Customer Service for a Fast Quote: 800-225-1600.
| Part Number | 06-54-0002 |
|---|---|
| Country of Origin | CR |
| UNSPSC Category | 42222001 |
| Harmonization Code | 9023.00.0000 |
| Mfg# | 40002 |
| Warranty | No Warranty |